Product Configurator

Ambulance Oxygen Delivery System for Two Cylinders

Automatic Source Changeover System with Intelligent Touch Panel Display

The Ambulance Oxygen Delivery System for Two Cylinders is a vital component for emergency medical services, enabling a continuous flow of oxygen to patients in need during transport to medical facilities.

Supply System for Two Cylinders

The System manages two sources and provides a continuous supply and smooth automatic switchover between the sources.
A touchscreen alarm panel monitors the gas pressure in each cylinder and shows/alerts the medical staff in case of a pressure drop.
Oxygen outlets enable easy and safe connection to the system, ensuring a constant and reliable oxygen flow for the patient.

Watch our video to discover more about the Ambulance Oxygen Delivery System for Two Cylinders

Request Information

Highlights

  • Automatic changeover between cylinders to ensure continuous and reliable oxygen flow.
  • Medical Gas outlet type: Sil-Ohio, ENV 737-6, Puritan Bennett type, DIN, British standard, DISS*.
  • Control Alarm panel monitor highlights:
    • Presenting the actual time left for treatment (in minutes) based on preset flow rate parameters.
    • Visual & audio alarm signals in case of low pressure in the change between cylinders.
    • 10.1” full HD LCD touchscreen – graphically and digitally displaying the cylinder gas pressure and output pressure, alerting in case of a pressure drop.
    • Graphically & digitally display of the actual status of each source – “In use” or “Stand by”.
    • Periodical maintenance alert.
    • Advanced Communication: TCP/ IP Modbus.
    • Gas expiration date alert – an alert for the medical gas cylinder expiration date.
    • Leak detection from cylinders.
    • Log History File- Event logging, recording all alarms and panel activities.
    • Customizable text & language.
    • A color trend arc showing the gas pressure status on the main line and at each cylinder.
    • Automatic adjustment of screen brightness.
  • Web remote view on the Internet – online view accessible via a web browser. This website allows authorized users to remotely access and view the data in real-time – optional.
  • Additional data: GPS data showing the current location of each ambulance by its number – optional.
  • SMS alert when oxygen pressure reduces to a pre-set level – optional.
  • 4G modem installed – optional.

 

Applicable Standards: 

  • ISO 10524, EN-1789, NFPA 1917

Silbermann’s Quality Management System: 

  • ISO 9001-2015
  • ISO 13485-2016

 

* Other types of oxygen gas outlets are available according to
customers’ requests.

Specifications

  • Gas sensors/Pressure 4-20 mA transmitters.
  • Oxygen Pressure Regulators**:
    • Input pressure: up to 200 bar
    • Output pressure: 3.5-4 bar (Regulated pressure)
    • Flow rate capacity of up to 100 liters per minute.
  • Automatic Changeover unit including manual bypass.
  • Flow selector 0-25 lpm
  • Power Requirements:
    • Voltage: 12V DC supplied from the Ambulance.
    • Electrical Supply Power.
    • 5A Slow Blow fuse 5x20mm.

 

** Oxygen regulator with CGA 540 connector – optional

FAQ

What is the Silbermann Ambulance Oxygen Delivery System for Two Cylinders?
How does the automatic cylinder changeover function work?
Can the system estimate the remaining oxygen treatment time?
Can the system detect oxygen leaks from cylinders?
Does the system keep a history of alarms and events?
What advanced monitoring and communication features does the system provide?
Can the system be monitored remotely over the Internet?
Can the system send SMS alerts when oxygen pressure is low?
Is the system suitable for ambulance manufacturers and vehicle integrators?
Why choose Silbermann’s Ambulance Oxygen Delivery System for Two Cylinders?

Contact Us

    We'd love to hear from you!

    #5 Harakevet st. POB 4605, Petah-Tikva ISRAEL
    +972-3-9309186 info@silbermnn.com