Product Configurator

DISS Type

DISS Standard Medical Gas Outlets for Safe Healthcare Applications

Silbermann designs and manufactures DISS-type medical gas outlets, also known as terminal units, to comply with standards:

  • NFPA 99
  • ISO 9170-1
  • CGA-V5

They are durable and safely connected to medical equipment, ensuring constant and reliable gas flow for patients.

Suitable for various hospital applications like anesthesia, ICU, and respiratory therapy.

DISS-type outlets are designed by industrial designers, helping to create a pleasant and comfortable environment for both patients and medical staff.

We thoroughly test each outlet for flow and leakage, and clean it for oxygen service before shipping.

Compatible with a range of gas types, including: Oxygen, Medical Air, and Vacuum.

It feature color-coded identification for improved safety and easy recognition.

Engineered for flexibility, the outlets can be installed on walls, panels, or pendants to meet the needs of various clinical settings

Request Information

Features and benefits

  • Accepts DISS gas-specific adapters.
  • All DISS outlets are 100% hydrostatic and tested for leaks. The outlets are factory-assembled, degreased, and cleaned for medical gas service.
  • The outlet uses gas-specific indexing to help prevent incorrect connections between medical gas services
  • 360-degree swivel inlet pipe for ease of installation.
  • Safe plug-in & release allowing parking of the male adapter on the outlet thread without pressure applied.
  • VITON seals in the front assembly help reduce maintenance needs and support reliable operation.
  • Modular singles – designed to provide ganging of rough-in modules in the field to form multiple at any order with 120 mm spacing.
  • Medical Gas System Standards:
    • NFPA99, EN ISO 7396-1 & G0-1 requirements.
  • Applicable Standards:
    • NFPA99, ISO 9170-1, CGA-V5.
  • Silbermann’s Quality Management System:
    • EN ISO 13485-2016
    • ISO 9001-2015

Technical Data

  • The medical gas outlet shall be SILBERMANN DISS type manufactured according to NFPA99.
  • The outlet shall be gas-specific by the DISS system.
  • The positive and negative pressure outlet shall consist of primary and secondary check valves to allow testing and online maintenance. The outlet shall be manufactured with 175 mm long, type K, 12.7 mm (1/2”) OD copper pipe. Rough-in pipe shall be silver brazed to the outlet body.
  • With the rough-in mounted, the outlet shall be depth, adjustable to compensate for various wall thicknesses.
  • The quick-connect outlet shall comply with ISO 9170-1 & CGA-V5 standard types and accept only the corresponding DISS Gas Specific Adapter.
  • Pressure drop of less than 20 kPa through the outlet at 120 LPM and 350 Kpa (3.5 Bar) inlet pressure. When fully open, the outlet allows a flow of up to 600LPM at a pressure of 400 KPa (4 Bars).

Contact Us

    We'd love to hear from you!

    #5 Harakevet st. POB 4605, Petah-Tikva ISRAEL
    +972-3-9309186 info@silbermnn.com